5-100 Nodal DOF Commands
SR780 Network Signal Analyzer
Interface Commands
*
RST
The *RST command resets the SR780 to its default configurations. The
communications setup is not changed. All other modes and settings are set to
their default conditions and values. This command takes some time to complete.
Do not send other commands on the same command line after *RST. Do not send
another command until *RST is complete. Either wait a fixed amount of time
(approximately 12 seconds) or use serial polling to wait for IFC to be set again in
the Serial Poll status before sending more commands.
*
IDN ?
The *IDN ? query returns the SR780's device identification string. This string is
in the format "Stanford_Research_Systems,SR780, s/n00001, ver007". In this
example, the serial number is 00001 and the firmware version is 007.
LOCL (?) {i}
The LOCL command sets the local/remote state. The parameter i selects LOCAL
(0), REMOTE (1) or LOCAL LOCKOUT (2). The states duplicate the GPIB
local/remote states. In the LOCAL state, both command execution and keyboard
input are allowed. In the REMOTE state command execution is allowed but the
keyboard and knob are locked out except for the [Help/Local] key which returns
the SR780 to the LOCAL state. In the LOCAL LOCKOUT state all front panel
operation is locked out, including the [Help/Local] key.
OVRM (?) {i}
The OVRM command sets (queries) the GPIB Overide Remote mode. The
parameter i selects No (0) or Yes (1). When Overide Remote is set to Yes, the
front panel is not locked out when the unit is in the REMOTE state.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......