Analog Input Menu 4-79
SR780 Network Signal Analyzer
The two displays have their own Engineering Units mode. This entry
field can be linked to both displays using the [Link] key.
Command: EU1M (?) {i} or EU2M (?) {i}
EU Label
Choose the Engineering Units Label for the Ch1 or Ch2 input. The
displayed units change from Volts to the EU Label when Engineering
Units are On. Choose USER for the user defined User Label.
Input units of acceleration (m/s
2
or in/s
2
), velocity (m/s or in/s), and
displacement (m or in) may be converted within a frequency domain
display using Transducer Convert in the [Display Options] menu. For
example, the FFT of an input configured for an accelerometer with
Engineering Units of m/s
2
may be displayed in displacement (meters)
using Transducer Convert.
The two displays have their own EU Label. This entry field can be linked
to both displays using the [Link] key.
Command: EU1L (?) {i} or EU2L (?) {i}
EU / Volt
Set the Engineering Units per Volt scale factor for the Ch1 or Ch2 input.
This factor is used to scale the display units from Volts to Engineering
Units when Engineering Units are On. The displayed units change from
Volts to the EU Label.
The EU scale factor may be entered in EU/V, V/EU, or dB(EU/V). In all
cases, the EU scale factor is displayed in EU/V along the vertical axis
whenever a measurement takes its input from an input with Engineering
Units On.
The two displays have their own EU scale factor. This entry field can be
linked to both displays using the [Link] key.
Command: EU1V (?) {x} or EU2V (?) {x}
EUs at Marker
Calculate the EU/Volt scale factor for either the Ch1 or Ch2 input based
upon the marker reading in the active display. Enter the desired
calibrated value for the marker reading (in the display units).
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......