3-12 Screen
Display
SR780 Network Signal Analyzer
Figure Chapter 3 -4 Single Display Screen
The Single display format is shown above. The measurement of the inactive display
continues even when it is not shown.
Data Graph
Data is normally graphed with signal amplitude on the Y axis and frequency or time on
the X axis. The physical size of the graph remains constant while the vertical and
horizontal scales may be changed. The graph area has a dotted grid for reference. There
are 10 horizontal divisions and either 8, 10, 12 or 15 vertical divisions. The horizontal
span is determined by the measurement span or time. The display normally shows the
entire measurement span. The X axis may be expanded (Zoom) and translated (Pan) to
display less than the full measurement span. This expansion does not change the
measurement span or time, it merely changes the display of the data.
Half Intensity Data Graph
Unsettled measurement data is graphed in half intensity. Once the measurement becomes
settled, the graph returns to full intensity.
FFT and Octave measurements are unsettled when the frequency span or resolution is
changed and when the input signal path is changed (gain, filters, etc.). The measurement
is not settled again until the discontinuity in the input data has propagated through the
digital filters and, for FFT’s, a complete new time record has been acquired.
FFT measurements running with a Time Record Increment less than 100% (overlapped
time records) display unsettled measurements in half intensity.
Octave measurements show unsettled octave bins in half intensity. Each octave bin is
displayed in half intensity until it is settled. The settling time of each bin is related to the
1/center frequency and the lowest bins take the longest time to settle.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......