1-64 Exceedance
Statistics
Use the knob to select (L) and press [Enter].
Choose L (L
eq
) as the sound level bin. L
eq
is a
standard broad band sound measurement. The result
is displayed as the last bin in the display and is
labeled ‘L’.
Press [Auto Scale A]
Scale DisplayA to show the entire range of the data.
6. Press [Waterfall]
Select the Waterfall menu. L
n
is calculated from
measurements stored in the waterfall buffer.
Press
<Storage>
Select waterfall Storage.
Use the knob to select (One Shot) and press
[Enter].
Choose One Shot to fill the waterfall buffer once
and stop.
Press <Total Count>
Change the number of measurements to store in the
waterfall buffer.
Press [1] [0] [0] [Enter]
Enter the Total Count for both displays (100).
The Storage Interval is 100 ms. A measurement
snapshot is added to the waterfall buffer every
Storage Interval so 100 measurements will take 10 s
to complete.
Press [Start/Reset]
Start the measurement. This resets the waterfall
buffer. New measurements are added to the
waterfall buffer every 100 ms. The number of
records stored in the waterfall is shown in the
Vertical Scale Bar and increments to 100.
Because the waterfall storage is One Shot, the
waterfall buffer fills once (100 records). No more
records are added after 10 seconds.
7. Press [Analysis]
Select the Analysis menu.
Press <Exceedance Stats>
Select the Exceedance Statistics menu.
Press <Stop Index>
The exceedance is calculated using the records in
the waterfall buffer starting with the Start Index (0
is the most recent record) and continuing through
the Stop Index (Total Count - 1).
Press [9] [9] [Enter].
Enter 99 to include records 0 through 99 (100 total).
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......