Waterfall
Commands
5-65
SR780 Network Signal Analyzer
Waterfall Commands
WDSP (?) d {, i}
The WDSP command sets (queries) the Waterfall Display Mode for display d.
The parameter i selects Normal (0) or Waterfall (1).
This command is valid only when the Measurement Group is FFT or Octave.
WSTO (?) d {, i}
The WSTO command sets (queries) the Waterfall Storage for display d. The
parameter i selects Off (0) or On (1).
This command is valid only when the Measurement Group is FFT or Octave.
WTOT (?) d {, i}
The WTOT command sets (queries) the Waterfall Total Count for display d. The
parameter i is a number of records. If the allocated memory is too small for the
requested number of records, an error occurs.
This command is valid only when the Measurement Group is FFT or Octave.
WAVA ? d
The WAVA command queries the number of records available in the waterfall
buffer for display d.
Records are not stored in the buffer unless Waterfall Storage is on (Continuous or
One Shot).
This command is valid only when the Measurement Group is FFT or Octave.
WFSK (?) d {, i}
The WFSK command sets (queries) the Waterfall Skip count for FFT
measurements for display d. The parameter i is a number of records.
This command is valid only when the Measurement Group is FFT.
WOSK (?) d {, x}
The WOSK command sets (queries) the Waterfall Skip count for Octave
measurements for display d. The parameter x is a time from .008 to 1000 s.
This command is valid only when the Measurement Group is Octave.
WVCT (?) d {, i}
The WVCT command sets (queries) the Waterfall View Count for display d. The
parameter i is a number of records.
This command is valid only when the Measurement Group is FFT or Octave.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......