Window
Menu
4-117
SR780 Network Signal Analyzer
Correlation Windows:
When the measurement is Auto Correlation, Cross Correlation, Windowed Time or a
User Function, the available windows are:
[-T/2..T/2]
This is a uniform window which should only be used on data which is self windowing
(lasts less than half of the time record).
[0..T/2]
This window zeroes the second half of the time record.
[-T/4..T/4]
This window zeroes the first and last quarter of the time record.
Octave and Swept Sine measurements do not use windows.
Force Length
Set the Force Window Length as a percentage of the FFT time record [1%..100%]. Points
in the time record up to the Force Length are unmodified. Points in the time record past
the Force Length are zeroed.
Choose Force as the FFT window before setting the Force Length.
The two displays can have different Force Lengths. This entry field can be linked to both
displays by using the [Link] key.
Command: FWFL (?) d {, i}
Expo TC
Set the Expo Window Time Constant as a percentage of the FFT time record
[5%..1000%]. This is the point where the window function reaches 1/e.
Choose Expo as the FFT window before setting the Expo Time Constant.
The two displays can have different Expo Time Constants. This entry field can be linked
to both displays by using the [Link] key.
Command: FWTC (?) d {, i}
Trace to Window
Copy a stored data trace (real part only) to the User window function of the active
display. Use the knob to pick a trace which already contains data and press [Enter]. The
window automatically switches to User window.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......