3-32 Keypad
SR780 Network Signal Analyzer
Function Keys
[Auto Scale A]
[Auto Scale A] automatically sets the vertical scale of DisplayA to show the entire range
of the data. Horizontal scaling is not affected.
Auto scale only operates on the data which is displayed on the graph. If the graph is
expanded, data corresponding to frequency or time bins which are not shown do not
figure in the auto scaling calculations.
[Link] [Auto Scale A] auto scales DisplayA and sets the vertical scale of DisplayB to
match DisplayA.
Command: ASCL d
[Auto Scale B]
[Auto Scale B] automatically sets the vertical scale of DisplayB to show the entire range
of the data. Horizontal scaling is not affected.
Auto scale only operates on the data which is displayed on the graph. If the graph is
expanded, data corresponding to frequency or time bins which are not shown do not
figure in the auto scaling calculations.
[Link] [Auto Scale B] auto scales DisplayB and sets the vertical scale of DisplayA to
match DisplayB.
Command: ASCL d
[Auto Range Ch1]
[Auto Range Ch1] toggles Channel 1 Input Ranging between Manual and Auto. In
Manual Ranging, the Input Range is set within the [Input] menu. In Auto Range, the
Input Range is adjusted automatically according to the Ch1 AutoRange Mode (also in the
[Input] menu). In Normal Auto Range, only overloads cause the range to change. In
Tracking Auto Range, the Input Range moves up for overloads and down when the signal
falls below half scale.
In Swept Sine group, Auto Range is always tracking.
The Ch1 Input Range indicator at the top of the screen will be shown in inverse if Ch1 is
Auto Ranging.
Auto Range responds to all frequencies present at the input (except those attenuated by
AC coupling), not just those within the measurement span.
[Link] [Auto Range Ch1] toggles the Input Ranging of both channels.
Command: A1RG (?) {i}
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......