Trigger
Menu
4-85
SR780 Network Signal Analyzer
Trigger Menu
The Trigger menu configures the Trigger Mode, Source, Level and Delay.
Watch Out For Triggered Sources
If a triggered source is selected (Chirp, Burst Chirp, Burst Noise or Arbitrary), the
triggered FFT measurement phase is stable only if the input signals are derived from the
source output. Turn the source off (or set it to Sine) when making triggered
measurements without the source.
Trigger Mode
Select the Trigger Mode [Free Run, Auto Arm, Manual Arm].
Free Run lets the measurements without a trigger. Measurements begin with [Start/Reset]
and run continuously or until they are complete.
Auto Arm enables triggering whenever measurements are running (with [Start/Reset]).
For FFT measurements, auto arming means that as soon as both time records are
processed, the trigger is immediately re-armed. Time records are taken as fast as the
trigger delay and time record processing permit. Two FFT measurements are always
triggered together.
Manual Arm enables triggering after [Start/Reset] only after the trigger is armed using
<Manual Arm> or from an interface command. Once the trigger is armed, the next trigger
event will trigger the measurement. For FFT measurements, each time record requires the
trigger to be re-armed.
Triggers are ignored unless measurements are started (with [Start/Reset]) and the trigger
is armed. Triggers are ignored while paused.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......