Error and system messages
11
The analyzer can detect and display various fault statuses. These fault statuses are divided
into maintenance requests and faults.
11.1
Maintenance requests
Maintenance requests are references to certain changes in the analyzer which - at the time
of occurrence - have no influence on the analyzer measurements. However, remedial
measures are recommended to guarantee that measurements remain possible.
If the analyzer is in measuring mode, you can recognize the occurrence of a maintenance
request in that an "M" appears at the right edge of the measurement screen.
Maintenance requests are logged and can be called in input mode
using the menu path "Analyzer status - Status - Maintenance re-
quests" (see section Analyzer status: Status: Maintenance request
(Page 105)). The corresponding message texts are stored. You can
delete the messages by pressing the <ENTER> key. However, they
appear again if the cause has not been eliminated.
The analyzer outputs a maintenance request in the following cases:
●
Message "AUTOCAL drift beyond tolerance"
The zero drift of a component is too large during the AUTOCAL procedure. The
AUTOCAL parameters can be entered as described in section Calibration:
AUTOCAL/drift values (Page 122). Using the conditions defined for the AUTOCAL
deviation (see section Configuration: Special functions: AUTOCAL deviation (Page 144)),
the actual deviation may deviate from the maximum permissible value. In such cases it
may be meaningful to set a shorter interval between two AUTOCAL procedures. If this
does not improve the situation, contact the servicing department.
●
Message "O2 sensor"
The measured voltage of the O
2
sensor has dropped as a result of aging, but is still within
the permissible range. This means that it is not necessary to replace the O
2
sensor
immediately, but this will soon be exhausted. This would be the correct time to order a
new O
2
sensor.
●
Message "Temperature beyond tolerance"
The contrast control is no longer guaranteed if the LCD temperature is outside the
permissible tolerance. It may then be difficult to read the display, or it could remain dark
in the worst case. If this fault occurs because of an excessively high ambient
temperature, provide sufficient ventilation or air conditioning. If the fault still occurs,
contact the servicing department.
●
Message "H
2
S sensor status"
If this message occurs, the measuring reserve of the H
2
S sensor is almost used up. We
recommend that you then replace the sensor. If the measuring reserve of the H
2
S sensor
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...