Error and system messages
11.2 Faults
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
179
Message
Possible causes
Remedy
Pressure of ambient air beyond toler-
ance
Pressure sensor faulty
Flow too low during measuring
Sample gas path blocked or leaky
Clean or replace the blocked parts
(hose, filter etc.). If the problem is still
present: Inform the service department
Pump not running
Start the pump as described in section
Configuration: Inputs/outputs/pump
(Page 134)
Pump capacity too low
Increase the pump capacity as de-
scribed in section Parameters:
Pump/LCD contrast: Pump (Page 130)
Pump faulty
Pump must be replaced. Inform the
service department
No data for temperature compensa-
tion
Temperature compensation not com-
pleted successfully
New components have been loaded
EEPROM has been initialized
Download the factory data as described
in section Configuration: Special func-
tions: Factory data/reset/units: Load
factory data (Page 149)
Flow too low during AUTOCAL
Sample gas path blocked or leaky
Clean or replace the blocked parts
(hose, filter etc.). If the problem is still
present: Inform the service department
Pump not running
Start the pump as described in section
Configuration: Inputs/outputs/pump:
Pump at CAL/MEAS (Page 142)
Pump capacity too low
Increase the pump capacity as de-
scribed in section Parameters:
Pump/LCD contrast: Pump (Page 130)
Pump faulty
Pump must be replaced. Inform the
service department
Concentration of O
2
too low
Measured value display: *****
O
2
sensor faulty or no longer usable as
result of aging
Replace the O
2
sensor as described in
section Replacing the electrochemical
oxygen sensor (Page 171)
O
2
sensor zero not calibrated
Calibrate the zero point of the O
2
sensor
as described in section Calibration:
Electrochemical oxygen measuring
Fault at analog output
Output component could not be initial-
ized when switching on
Limits were violated when calibrating the
analog section
General fault of all IR channels
Measured value display: '*****'
Chopper faulty
Fault of addresses for IR channels
Plug-in jumper on detector for detection
of components not OK
The cable of the detector has no contact Check that the plug is correctly con-
nected to the detector (the plug must
latch in twice).
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...