Description
3.4 Technical specifications
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
33
Gas inlet conditions
Sample gas pressure
•
Without pump
•
With pump
Unpressurized (< 1200 hPa (17.4 psi) absolute)
Unpressurized suction mode, set in factory with 2 m (6 1/2 ft) hose
at sample gas outlet; full-scale value calibration necessary under
different venting conditions
Sample gas flow
72 ... 120 l/h (1.2 .. 2 l/min)
Sample gas temperature
0 ... 50 °C (32 ... 122 °F)
Sample gas humidity
< 90 % RH (relative humidity), non-condensing
Note
Since measuring ranges can be changed, all accuracy data applies to the ranges specified
on the label!