Commissioning
6.4 Commissioning
ULTRAMAT 23
82
Manual, 3/2016, A5E37100388-003
6.4
Commissioning
6.4.1
Commissioning
Once all preparatory work for commissioning has been completed, go through the following
checklist:
●
The analyzer is set to the correct operating voltage
●
All gas preparation elements are connected and ready for operation, and have been
checked for leaks
●
All required connections to and from the analyzer have been established
Following successful checking, connect the analyzer to the power supply and switch it on.
Wait for the warm-up phase to elapse (see Warm-up phase (Page 95)).
6.4.2
AUTOCAL
The analyzer carries out an automatic calibration with the connected medium during the
warm-up phase following switching-on. This AUTOCAL adjusts the zero point and sensitivity
of the IR channels. If an O
2
sensor is present, its sensitivity is additionally calibrated using
the ambient air (20.95% O
2
).
Note
Analyzers with H
2
S sensor
The hydrogen sulfide sensor is not calibrated during the course of this first AUTOCAL. The
zero point of the H
2
S sensor is only calibrated starting from the second AUTOCAL of the
analyzer.
Note
Analyzers without electrochemical O
2
sensor
In the case of analyzers without an electrochemical O
2
sensor, the AUTOCAL can be carried
out with nitrogen, but in the case of analyzers with an electrochemical O
2
sensor, it is
essential to use air. The correct medium is selected depending on the used configuration
(gas connections) and cannot be parameterized using the software.
Note
Analyzers with paramagnetic O
2
sensor
In the case of analyzers with a paramagnetic O
2
sensor, the input menu can be used to
select whether the AUTOCAL is to be carried out with air or N
2
, and thus whether the
sensitivity (20.95 % O
2
) or the zero point of the sensor is calibrated.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...