Functions
8.2 Calibration
ULTRAMAT 23
116
Manual, 3/2016, A5E37100388-003
The analyzer then compares the setpoint and actual value
(measured value) of the calibration. If the deviation between
the values is within the tolerance, the adjacent message 'o.k.'
is output.
If the actual value deviates by more than approx. 20% from the
setpoint determined in the factory, the message 'not o.k.' ap-
pears instead (no calibration possible). The causes of this
message can include:
•
Incorrectly entered setpoint
•
Calibration gas concentration does not agree with entered
value
•
The flow of calibration gas is insufficient
By pressing the <ESC> key you can exit the calibration procedure.
8.2.4
Calibration: Paramagnetic oxygen sensor
In this menu you can call the following functions for the para-
magnetic oxygen sensor:
•
"AUTOCAL O2 sens."
–
YES (factory setting): The zero point or sensitivity of the
paramagnetic oxygen sensor is set with each
AUTOCAL. Selection of zero point or sensitivity is
carried out using the parameter in the 2nd line
'AUTOCAL with N2'.
–
NO: No calibration of the paramagnetic oxygen sensor
during an AUTOCAL.
•
"AUTOCAL with N2"
–
YES: The AUTOCAL is carried out with nitrogen, where
the zero point of the sensor is calibrated.
–
NO: (factory setting ) No calibration of the paramagnetic
oxygen sensor during an AUTOCAL
•
"Calib. O2 zero point"
This function is used to calibrate the zero point of the
paramagnetic sensor
•
"Calib. O2 range"
This function is used to calibrate the full-scale value or
sensitivity of the paramagnetic sensor and to set the
setpoint.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...