Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
149
Configuration: Special functions: ELAN/PROFIBUS/external interference: Cross-interference
The adjacent menu display appears when you select this
parameter. You can use this function to:
•
Switch the correction of cross-interference on or off for the
duration of the calibration.
To do this, you must select the 2nd line and switch the
parameter on or off there.
–
OFF (factory setting) means that the correction of
cross-interference is switched off during the calibration.
–
ON means that the corrections of cross-interference
remain active during the calibrations. It is thus possible
to use gas combinations as calibration gases.
•
Switch the correction of a constant cross-interference on
or off. To do this, you must select the 3rd line and switch
the parameter on or off there. If the constant cross-
interference is switched ON, the measured value of the
selected component is corrected with the entered value by
adding.
8.4.2.4
Configuration: Special functions: Factory data/reset/units
Configuration: Special functions: Factory data/reset/units
You can use this menu to select a number of items with which
e.g. inappropriate configurations and analyzer settings can be
cancelled:
The adjacent menu display appears when you select the func-
tion from the higher-level menu.
Configuration: Special functions: Factory data/reset/units: Load factory data
In this menu you can reestablish the original parameters pre-
sent when the analyzer was delivered.
Note:
All modifications (parameters and configuration) which you have
made since then are deleted.
The adjacent display appears when you select this function.
You define the further sequence by pressing either the
<ENTER> or <ESC> key.
When you select this function, the adjacent display appears for
the duration of the load procedure.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...