Description
3.4 Technical specifications
ULTRAMAT 23
36
Manual, 3/2016, A5E37100388-003
Influencing variables
Drift
•
With AUTOCAL
•
Without AUTOCAL
Negligible
< 2 % of smallest measuring range/week
Temperature
max. 2 % of smallest possible measuring range according to label per
10 K with an AUTOCAL cycle time of 6 h
Air pressure
< 0.2 % of measuring range per 1 % change in pressure
Power supply
< 0.1 % of output signal span with a variation of ±10 %
Line frequency
± 2 % of full-scale value with a frequency variation
of ± 5 %
Deviations with measuring range 0 ... 200 mg/m
3
SO
2
(analyzer versions 7MB2335-xNBxx-xAAx, 7MB2337-xNBxx, 7MB2337-xxxxx-xNBx, 7MB2338-xxxxx-xNBx)
Availability
Max. 95 %
AUTOCAL cycle time
Max. 6 h
Temperature variations
Max. 1 °C (1.8 °F)
The device must not be operated in an area subject to drafts. This is
especially valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.
Deviations with measuring range 0 ... 100 mg/m
3
SO
2
(analyzer versions 7MB2335-xNTxx-xAAx, 7MB2337-xNTxx-xxxx, 7MB2337-xxxxx-xNTx, 7MB2338-xxxxx-xNTx)
Delayed display (T
90
time)
Depending on the external gas conditioning.
Note: SO
2
is highly soluble in water
Output signal resolution
≤ 0.1
mg/m
3
Linearity error
≤ 2 mg/m³ (2% of smallest measuring range)
Drift
•
With constant ambient conditions
•
After temperature jump
≤ 0.33 mg/m³ per hour after a running
-in period of 200 hours
Not defined
Temperature influence
≤ 8
mg/m³ per 10 K dev
iation; additional drift after temperature jump
Influence of auxiliary power
≤
0.4 mg/m³ with a change of ±
10%
Influence of line frequency
≤
8 mg/m³ with a change of ±
5%
Cross-sensitivities
Refer to following table
Availability
Max. 95 %
AUTOCAL cycle time
Max. 3 h
Temperature variations
Max. 1 °C (1.8 °F)
The device must be operated at a constant ambient temperature. The
device must not be operated in an area subject to drafts. This is especial-
ly valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...