Functions
8.1 Diagnostics
ULTRAMAT 23
104
Manual, 3/2016, A5E37100388-003
8.1.1
Diagnostics: Status
In this menu you can call all status messages of the
ULTRAMAT 23 via further submenu items.
In this example, the status of the O
2
sensor is shown in the last
line. If the analyzer is equipped with software for operating an
H
2
S sensor, the text 'Probe status' is shown in the last line with
the following options:
•
Only H
2
S sensor:
The analyzer calls this function directly.
•
H
2
S and O
2
sensors:
You will be requested to select the associated sensor.
8.1.1.1
Analyzer status: Status: Logbook/faults
This dialog displays the contents of a logbook. This con-
tains all recorded faults. Each type of fault only appears
once in the logbook and is output in alphanumeric text (an
overview of the possible faults can be found in section
Faults (Page 178)).
Viewing the logbook is protected by code level 1.
Following access to the logbook you can:
•
Display all recorded faults in succession using the
<→>
key
•
Delete the currently displayed fault message using the
<ENTER> key.
If further faults are present, these appear in succession.
You should delete all fault messages whose causes
have been eliminated.
Note
By deleting the fault message you do not eliminate the
cause of the fault (see section Faults (Page 178)).
A corresponding text is output when all stored faults have
been displayed. Terminate display of the logbook using the
<
→>
key.
Note
If the analyzer is in measuring mode, you can recognize
the occurrence of a fault in that an "F" appears at the right
edge. A "! at the right edge signals that a fault has been
logged which is no longer present.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...