Functions
8.2 Calibration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
119
8.2.5.1
Calibration: H2S sensor: Defining the installation
You have to re-enter the date of installation after installing a new sensor.
The adjacent display appears when you select the 'Installation
date' menu item:
Now you can enter the date of installation in the form:
'DD.MM.YYYY'.
Note
Prior to entering the installation date, you must perform a zero point and measuring range
calibration of the H
2
S sensor, otherwise the new date will not be accepted.
8.2.5.2
Calibration: H2S sensor: Calibrating the zero point
You can use this function to re-adjust the zero point of the H
2
S sensor. As the zero gas you
can use:
●
Nitrogen
or
●
air free of H
2
S.
To do this, navigate as follows: Calibration -> Please enter code -> Calibrate probe ->
Choose component -> H
2
S -> Calibrate zero.
The adjacent display appears.
In the case of analyzers with an internal sample gas pump,
selecting this function switches from the sample gas inlet (inlet
1) to the zero gas inlet (inlet 3).
Calibrate the zero point as follows:
•
Connect zero gas to the analyzer and observe the display.
•
Wait until the displayed value has stabilized.
•
Commence with correction of the zero point by pressing the
<ENTER> key.
•
Exit the calibration by pressing the <ESC> key.
The message "Tolerance not o.k." appears if an error occurs during the calibration.
8.2.5.3
Calibration: H2S sensor: Calibrating the measuring range
You can enter the setpoint and calibrate the sensitivity of the sensor using this function.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...