Service and maintenance
10.2 Maintenance work
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
175
Proceed as follows to install the new sensor:
1.
Tighten the new H
2
S sensor by hand in the holder up to the mechanical limit (shaded)
Figure 10-2 Sensor holder
①
(shaded)
2.
Plug the cable with the connector onto the sensor.
3.
Close the analyzer by screwing tight the cover.
4.
Switch the analyzer on and wait for the warming-up phase.
5.
Check for leaks in the device
This is described in section Leaks in the gas paths (Page 81)
6.
Calibrate the H
2
S sensor as described in section Calibration: H2S sensor (Page 118)
7.
Enter the installation date with the menu command "H
2
S Installation date".
The analyzer is then ready for use again.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...