Error and system messages
11.2 Faults
ULTRAMAT 23
178
Manual, 3/2016, A5E37100388-003
is used up completely, the fault message "Measured value channel 3 beyond tolerance"
is displayed. A measurement is no longer possible.
●
External maintenance requests
These are signaled via the binary inputs. The analyzer must be equipped with an add-on
board for this.
11.2
Faults
Fault messages are references to certain changes in the analyzer which influence its ability
to measure correctly. In such cases remedial measures are required.
If the analyzer is in measuring mode, you can recognize the occurrence of a fault in that an
"F" appears at the right edge of the measurement screen.
Faults are logged and can be called in input mode using the menu
path "Analyzer status - Status - Logbook/faults" (see section Diag-
nostics: Diagnostics values (Page 107)). The corresponding mes-
sage texts are stored as alphanumeric text in the logbook. You can
delete the messages by pressing the <ENTER> key. However,
they appear again if the cause has not been eliminated.
The following table provides a summary of fault messages, their causes, and measures to
eliminate the faults.
If nothing is specified for a fault message in the 'Remedy' column, you must contact the
servicing department when this message occurs.
Message
Possible causes
Remedy
Measured value channel 1 beyond
tolerance
Measured value display: *****
Analyzer unit of first component is faulty
Measured value channel 2 beyond
tolerance
Measured value display: *****
Analyzer unit of second component is
faulty
Measured value channel 3 beyond
tolerance
Measured value display: *****
Analyzer unit of third component is faulty
Measured value channel 3 (H
2
S
sensor) beyond tolerance
Measured value display: *****
Sensor faulty
Replace the H
2
S sensor as described in
section Replacing the hydrogen sulfide
Measured value O
2
beyond tolerance
Measured value display: *****
Electrochemical O
2
sensor faulty or no
longer usable as result of aging
Replace the electrochemical O
2
sensor
as described in section Replacing the
(Page 171).
Supply voltage beyond tolerance
Supply voltage varies
Correct such that the supply voltage
remains stable within the tolerance
values permissible for the analyzer.
Power supply unit on motherboard faulty
Temperature of analyzer beyond
tolerance
Ambient temperature too high or too low Provide sufficient ventilation or air condi-
tioning.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...