Functions
8.2 Calibration
ULTRAMAT 23
120
Manual, 3/2016, A5E37100388-003
Navigate as follows to select this function: Calibration -> Please enter code -> Calibrate
sensor -> Select component -> H
2
S -> Calibrate MR.
The adjacent display appears.
Proceed as follows to enter the setpoint:
1.
Position the cursor on the 2nd line of the display (Setpoint
MR 1+2)
2.
Press the <ENTER> key
The adjacent display appears. You can now enter the setpoint
of the calibration gas:
Calibrate the sensitivity as follows:
1.
Position the cursor on the 3rd line of the display
(Start cal. MR 1+2).
2.
Press the <ENTER> key.
The adjacent display appears. Now carry out the following
steps for the sensitivity calibration:
1.
Connect the sample gas inlet to the calibration gas.
2.
Inject calibration gas to the sensor with a flow rate of
1.2 ... 2.0 l/min.
3.
Wait until the measured value has stabilized.
4.
Then press the <ENTER> key.
5.
Exit the calibration by pressing the <ESC> key.
The message "Tolerance not o.k." appears if an error occurs during the calibration.
8.2.5.4
Calibration: H2S sensor: Enter TC parameters
You use this function to enter the temperature compensation parameters for calibration of
the zero point and sensitivity. You can read these parameters from the sensor.
To do this, navigate as follows: Calibration -> Please enter code -> Calibrate probe ->
Choose component -> H
2
S -> Calibrate TC parameters. The following screen appears:
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...