Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
137
●
Zero:
See following table:
Selectable
Analog current
0 - 20 mA
0 mA
2 - 20 mA
2 mA
4 - 20 mA
4 mA
NAMUR - 20 mA
3 mA
●
21 mA:
see following table:
Selectable
Analog current
0 - 20 mA
21.0 mA
2 - 20 mA
21.0 mA
4 - 20 mA
21.0 mA
NAMUR - 20 mA
21.5 mA
Analog output during fault
Here, you can define the type of measured value output during
a fault.
Output of the following measured values is possible here:
●
Hold: The value measured directly prior to commencement of a fault is output unchanged.
This also applies to the limits which are output (see section Parameters: Limits
(Page 128)).
●
Actual: The measured value is continuously updated.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...