Functions
8.4 Configuration
ULTRAMAT 23
150
Manual, 3/2016, A5E37100388-003
Configuration: Special functions: Factory data/reset/units: Reset
You can use this function to restart the analyzer.. When you
select this function, you will be warned (see
adjacent display) that the analyzer initially runs through a warm-
up phase following the restart and is thus unavailable for meas-
urements for a certain time.
The <ENTER> key initiates the restart with the warm-up phase.
Triggering of the restart can be prevented here using the
<ESC> key.
Configuration: Special functions: Factory data/reset/units: Change units
In the second line of this menu display, you can change the
factory-set units of the measured components.
After changing the unit, the display of the 'MB min' and 'MB
max' parameters is adapted accordingly.
This dialog is specific to the component.
Note
The full-scale values may assume unusual values as a result of the component-specific
conversion factors. Subsequent adaptations can be carried out as described in
section Parameters: Measuring ranges: Setting measuring ranges (Page 126). In addition,
you should also check these parameters following this change:
•
Setpoints of the calibration gases (section Calibration: Infrared measuring range: Set
span gas values (Page 113))
•
Settings for the limits (section Parameters: Limits (Page 128)).
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...