Operation
7.4 Operating modes
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
101
Once you have carried out the above sequence on the analyzer, you are already acquainted
with the important points for operation of the ULTRAMAT 23.
7.4.3.3
The ESC key
You can trigger two different functions by pressing the <ESC> key:
●
Firstly, you can cancel a commenced procedure, e.g.:
–
The input of a number
–
A calibration procedure with calibration gas
–
Any function if a fault occurs, e.g. if the flow of sample gas to the analyzer is missing.
●
Secondly, you can use the <ESC> key to move to the next higher level in menu structure
("scroll back"). This procedure is the opposite to selection of a submenu using the
<ENTER> key ("scroll forwards"). If you repeatedly press the <ESC> key, you return back
to the main menu step-by-step. If you press the <ESC> key again in the main menu, the
analyzer switches over from input to measuring mode. All inputs are imported at the
same time. However, you will not be asked to confirm the inputs.
An example will clarify this:
The analyzer is in measuring mode and is uncoded.
Switch from measuring mode to input mode using
<ENTER>, select the menu item "Parameters" using the
<
↑
> or <
↓
> key, and press <ENTER> to confirm.
In this manner, you enter the first submenu.
Now press <ESC> and then
<ENTER> again. You have returned by one level and then
moved forwards by one level again; you are therefore in the
same menu again.
Press the <ESC> key twice, you are back in measuring
mode again.
7.4.3.4
The CAL key
If the analyzer is in measuring mode, pressing the <CAL> key triggers a single, automatic
calibration with ambient air or nitrogen (AUTOCAL).
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...