Service and maintenance
10.2 Maintenance work
ULTRAMAT 23
172
Manual, 3/2016, A5E37100388-003
10.2.8
Replacing the hydrogen sulfide sensor
DANGER
Danger of poisoning
The replacement of the sensor module represents interference in the gas path. The sample
gas circulating therein may contain toxic components that lead to death in certain
concentrations.
To ensure that the sample gas path is free of toxic material when replacing the sensor
module, the gas path must be flushed with ambient air or nitrogen for a period of about 10
minutes before performing the task.
WARNING
Danger of electric shock
The device will be open when the sensor is being replaced. This means present dangerous
contact voltage will be present, which can lead to electric shock.
For this reason, the sensor module may only be replaced when power is off.
WARNING
Danger of chemical burns
The H
2
S sensor contains sulfuric acid, which leads to burns on unprotected skin.
Therefore do not use any tools when replacing the sensor module which could damage the
sensor due to sharp edges or squeezing.
If contact with the acid occurs nevertheless, rinse the affected skin immediately with plenty
of water!
NOTICE
Improper disposal
The exhausted or faulty H
2
sensor is hazardous waste and must be packed and disposed
of accordingly!
The exhausted H
2
S sensor is electronic waste with the code number 160215, that is, a
"dangerous component removed from used devices". It must therefore be disposed of
correctly and in an environmentally-friendly manner by a local waste disposal company.
Environmental damage may occur if this stipulation is not followed. The polluter is also
threatened with criminal action!
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...