Functions
8.1 Diagnostics
ULTRAMAT 23
108
Manual, 3/2016, A5E37100388-003
•
Raw values are the measured values in % of the full-scale
value (= 100 %).
•
Not limited concen. are the measured values as they are
also displayed in measuring mode. You can approximately
display the concentrations here even with an upward or
downward violation of the largest range. Negative values
are also displayed (live zero). The current measuring range
is output on this display in the last two columns.
8.1.2.2
Analyzer status: Diagnostics values: (Electrochemical) O2 sensor
This dialog displays the diagnostics values of the electrochem-
ical oxygen sensor (option). Meaning:
•
Sensorsig is the current pressure-compensated voltage of
the O
2
sensor in mV
•
O
2
is the current oxygen value. Negative values are also
possible here
•
Inst. date is the installation date of the O
2
sensor (see
section Calibration: O2 measuring range: Sensor inst. date
(Page 114))
•
Inst.volt is the pressure-compensated voltage of the O
2
sensor when it was installed.
8.1.2.3
Analyzer status: Diagnostics values: (Paramagnetic) O2 sensor
This dialog displays the diagnostics values of the paramagnet-
ic oxygen sensor (option). Meaning:
•
Sensorsig is the current voltage of the O
2
sensor in mV
•
O
2
is the current oxygen value. Negative values are also
possible here
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...