Parameters: Measuring ranges: Setting measuring ranges
The full-scale values of the measuring ranges can be set in the
third and fourth lines of this dialog. They must lie within the
factory settings, i.e. if an analyzer is factory-set for a total
range from 0 to 2000 mg/m
3
, modifications are only possible
within this range. The following definitions also apply:
•
The smaller MR must not be greater than the higher MR.
•
The following input limits apply to the ranges:
–
Lower limit: 0.01 times the smaller MR according to
factory setting (label)
–
Upper limit: 1.1 times the higher MR according to factory
setting (label)
In the example:
•
Smallest MR 1: 0 to 4 mg/m
3
•
Highest MR 2: 4 to 2200 mg/m
3
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...