Functions
8.1 Diagnostics
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
105
8.1.1.2
Analyzer status: Status: Maintenance request
This dialog indicates the logged maintenance requests. A
maintenance request is set if the values of certain parameters
have reached defined limits, but the analyzer is still able to
measure (e.g. AUTOCAL deviation or O
2
sensor status; see
also section Analyzer status: Status: O2 sensor status
(Page 106)). A corresponding message is output in alphanu-
meric text.
Access is protected by code level 1.
Following access to the maintenance request list you can:
•
Display all recorded faults in succession using the <
→
> key
•
Delete the currently displayed fault message using the
<ENTER> key. If present, the next maintenance request is
then displayed. You should delete all maintenance requests
whose causes have been eliminated.
Note
If the analyzer is in measuring mode, you can recognize the
occurrence of a maintenance request in that an "M" appears at
the right edge.
8.1.1.3
Analyzer status: Status: AUTOCAL deviation
This dialog indicates the deviation in setpoint between several
AUTOCAL procedures. The parameters have the following
meanings:
•
The text in the two top lines provides information on the
number of AUTOCAL procedures which have been carried
out since the reference value for AUTOCAL was last set
(see section Configuration: Special functions: AUTOCAL
•
Deviation is the measured deviation of the actual value from
the reference value, displayed in % of the set measuring
range (with autoranging, range 1 is assumed). This
deviation must not be above the set maximum value.
•
Max. allowed is the maximum permissible value for the
deviation. Refer to section Configuration: Special functions:
AUTOCAL deviation (Page 144) for setting the maximum
value.
This function is specific to the component.
Summary of Contents for ULTRAMAT 23
Page 2: ......
Page 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Page 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Page 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Page 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Page 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Page 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Page 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Page 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Page 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Page 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...