Replacing boards / replacing components
6 - 33
Siemens AG
SPR2-230.061.01
Page 33 of 48
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
•
Configure the following parameters in User setup for MEMOSKOP
:
•
Perform the IQ quick test.
Roadmap K-Factor, phase B Program "MD 2".
Program a different value only upon customer request.
Roadmap K-Factor, phase C Program "MD 2".
Program a different value only upon customer request.
Roadmap Time of phase A
Program "4,5".
Program a different value only upon customer request.
Operating mode Parameters Change the programmed parameters only upon customer
request.
VCR output
Configure to monitor A.
Hardcopy Release
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors:
configure to both monitors.
Split screen
Configure horizontally.
Configure vertically only upon customer request.
Display of time
Configure Yes.
Set Date Time
Set the current date and time.
Hospital Name
Program the hospital name.
Storage of Dose on disk
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Dose on Hardcopy output
If area dose product meter is present, (Diamentor) con-
figure yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service instructions
LITHOSTAR MODULARIS.