Adjustments / Programs
5 - 21
Siemens AG
SPR2-230.061.01
Page 21 of 30
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Calibrating the area dose product measurement device
5
CONSTANT parameter (Qt * Kg) = T
w
5
•
Read and record the
T
w
value shown on the electronics unit of the area dose product
measurement device.
Programming
5
•
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
•
Start the service program.
•
Select the Adjustment, Parameter menu.
•
In the Combo box "Parameter Groups:", select "Diamentor" .
•
Click on "Get from Unit".
•
The Parameter "1. Diamentor Existing" must be on YES .
•
Select parameter "2. Constant (Qt * Kg)" .
•
In "Value Actual:"field, enter the parameter
T
w
.
•
Click on "Set Value".
•
Click on "Put to Unit" .
•
Click on "Save to File".
Fig. 6
This adjustment is used to calibrate the area dose pro
duct mea-
surement device of the SIREMOBIL Iso-C display.
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
Potentiometer
NOTE
NOTE