
SIREMOBIL Iso-C
SPR2-230.061.01
Page 16 of 48
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
6 - 16
Replacing boards / replacing components
Concluding work
6
•
Reinstall the I.I. on the C-arm flange. Refer to chapter 1, I.I. service position.
•
Check the collimator display on the monitor and, if necessary, adjust it.
•
Perform the I.Q. Quick test.
•
Check the setting of the laser light localizer, if present. If necessary, adjust it.
•
If the "MODULARIS" system is present, perform the isocenter check.
•
Execute the country-specific acceptances (§16 Partial acceptance test, DHHS,...).