SIREMOBIL Iso-C
SPR2-230.061.01
Page 6 of 10
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
7 - 6
Brakes / Movements
Vertical lift motor
7
•
Remove the vertical lift magnet from the vertical lift motor (three screws).
•
Remove the attachment screws for the vertical lift motor.
•
Remove the vertical lift motor with the gears from the spindle.
•
Disconnect the wires from the frequency converter FU.
•
Route the wires of the new vertical lift motor to the frequency converter FU and connect
them.
•
Place the new motor on the spindle. Make sure the spindle is seated correctly in the
gears.
•
Attach the motor with the screws.
•
Install the vertical lift magnet on the vertical lift motor.
•
Return the SIREMOBIL Iso-C to the upright position.
•
Lubricate the spindle and the lifting column mounting.
•
Check the vertical lift function. Also check that the vertical lift motor switches off at the
end positions of the vertical lifting column. (limit switches).
Fig. 6
Risk of injury!
Noncompliance can lead to medium to severe injuries and/or
property damage.
To replace the vertical lift motor, the SIREMOBIL Iso-C must be
placed on its side.
This requires 6 people. To prevent damage to the
SIREMOBIL Iso-C or to the floor, we recommend using several
wool blankets or corrugated cardboard, or other suitable floor
covering.
WARNING