Brakes / Movements
7 - 7
Siemens AG
SPR2-230.061.01
Page 7 of 10
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Vertical lift limit switches
7
•
Replace the limit switches. Reconnect the wires.
•
To adjust the limit switches, move the lifting column slowly and carefully by briefly tap-
ping the lifting column switch at the control console of the basic unit until it is positioned
just before the mechanical end stop.
•
Adjust the respective limit switch and tighten the screws.
•
Make sure the limit switch functions correctly.
•
To do this, move back out of the end position.
•
Then slowly and carefully move to the end position by briefly tapping the lifting column
switch on the control console of the basic unit.
•
The lifting column motor must switch off prior to reaching the end position and the lifting
column must come to a full stop.
•
Move the lifting column approximately 10 cm away from the end position.
•
Now move the lifting column at full speed in the direction of the end position by pressing
and holding the lifting column switch.
•
The lifting column motor must switch off prior to reaching the mechanical end stop and
the lifting column must come to a full stop.
Switch for blocking vertical lift
7
•
Remove the horizontal lift cover to replace both the switches. Refer to chapter 1 of these
instructions.
Adjustment: Refer to the MODULARIS service instructions.
Fig. 7