
SIREMOBIL Iso-C
SPR2-230.061.01
Page 12 of 48
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
6 - 12
Replacing boards / replacing components
•
Check the display of the blades on the monitor screen and adjust, if necessary.
•
Check collimation for direct exposure if there is a cassette holder.
•
Reinstall the rubber stop.
•
Ensure that the area/dose product measurement device is functioning properly
(if present).
•
Check the setting of the laser targeting device and, if necessary, adjust it according to
the Adjustment Instructions for the Laser Targeting Device, RXR2-130.032.01.
•
If the "MODULARIS" system is present, perform the isocenter check.
Replacing the I.I.
6
Checking the temperature indicator
6
•
Prior to installing the I.I., check the temperature indicator.
•
If the inner square field of the indicator is white, the I.I. did not exceed the temperature
range. Remove the temperature indicator.
•
If the indicator is discolored (inner field black), proceed according to the IQ document
RXD-000.038.01.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the SIREMOBIL Iso-C off and wait at least > 3 minutes until
the high voltage in the system and in the I.I. mini voltage supply
dissipates.
WARNING