SIREMOBIL Iso-C
SPR2-230.061.01
Page 28 of 30
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
5 - 28
Adjustments / Programs
Monitor zoom
Yes: When pressing the key for MEMOSKOP - Zoom, the
image will be displayed enlarged.
NO: The MEMOSKOP - Zoom key has no function.
For MEMOSKOP C-E set on NO. Otherwise set YES.
Monitor split
Yes: Horizontal or vertical image split possible.
No: No horizontal or vertical image split possible.
Program Yes.
Monitor text
Yes: Superimposes text.
No: No text superimposition.
For SIREMOBIL Iso-C always set on Yes.
Maximum no. of images
The number of stored images can be progemmed.
Program the value to 700 for MEMOSKOP C and to 900 for
MEMOSKOP SUB.
For MEMOSKOP C-E 100 program the value to 100
For MEMOSKOP C-E, this parameter is not used.
Default programs chan-
geable
Default: Yes
The organ programs can be protected against changes by
setting "No".
Default LUT
After switch-on, the programmed brightness/contrast curve
(LUT) is selected automatically: Default: LUT 2
Default LUT for Subtraction
After switch-on, the programmed brightness/contrast surve
(LUT) is selected automatically for the subtraction mode,
when equipped with a MEMOSKOP C-Sub.
Default: LUT 1
Default Edge
After switch-on, the programmed edge enhancement is
selected automatically: For MEMOSKOP SUB set on 1.
Software compatibility
Not implemented. Must be set to "No".
Unit serial number
Enter the serial number of the SIREMOBIL Iso-C when it is
equipped with a DICOM bridge. The serial number is requi-
red to identify the images transferred. Please note that you
have to omit the leading 0 when entering the serial number.
Disable dose rate selection
in user Setup
This means "Yes" for SIREMOBIL Iso-C without an area
dose product measurement device. This means "No" for a
SIREMOBIL Iso-C with an area dose product measurement
device.
Subtraction K-factor,
phase A
Noise suppression, phase A, is programmed in the
SUB-mode.
Phase A: Mask acquisition.
Subtraction K-factor,
phase B
Noise suppression, phase B, is programmed in the
SUB-mode.
Phase B: Acquisition of fill image.
Subtraction Time of phase A The time of phase A in the SUB-mode is programmed.
Phase A: Mask acquisition.
Roadmap K-factor, phase A
Noise suppression, phase A, is programmed in the road-
map mode.
Phase A: Mask acquisition.
Function:
Comments: