Adjustments / Programs
5 - 11
Siemens AG
SPR2-230.061.01
Page 11 of 30
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Collimator, X-Iris
5
•
Attach the center cross in the center of the image intensifier input screen.
•
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
•
Start the service program.
•
Select the "Adjustment, Calibration" menu.
•
Answer the query with "yes".
•
Select "X-Iris" in the "Function Groups" combo box.
•
In the window "Available Functions:" select "1. Init (Limits Off / Autosearch)" and click on
Execute.
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the Unit", click "No".
•
The X-Iris can now be completely opened and closed with the X-Iris keys.
•
Select I.I. survey format.
•
During fluoroscopy, adjust the X-Iris by opening and closing it with the X-Iris keys so that
the collimator blades are still visible along the edge of the I.I.
•
Select "2. II. - Full Format Position" on the service PC and click on "Execute" .
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the Unit", click "No".
•
Switch to zoom format.
•
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
•
Select "3. II. - Zoom Format Position" on the service PC and click on "Execute" .
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the Unit", click "No".
•
Open the X-Iris completely.
•
Select "4. CEX Position" on the service PC and click on "Execute" .
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the Unit", click "No" .
•
Close the X-Iris completely.
•
Release fluoroscopy. The X-Iris must close to < 4.5 cm .
Because of the mechanical hysteresis of the semi-transparent
slot diaphragm, open the collimator blades to maximum aperture,
prior to adjusting the radiation iris.
X-ray radiation!
See Chapter 1, Safety Information.
Observe the need for radiation protection. Wear a lead apron.
NOTE
WARNING