7 - 1
Siemens AG
SPR2-230.061.01
Page 1 of 10
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Brakes / Movements
7
Brake force / Brake torque
7
Measuring the horizontal force required for movement
7
The attachment point for the spring scale is the end of the railing near the image intensi-
fier in sector B and/or the handgrip of the horizontal carriage. The carriage must always
be pulled with a consistent speed.
¹
The maximum force for moving the horizontal carriage consistently when the
brakes are released is 80 N.
¹
When the brakes are applied, the force must be > 180 N.
Measuring the horizontal swivel force
7
¹
Swivel force when the brakes are released max. 35 Nm.
¹
Swivel force when brakes are applied > 100 Nm.
Fig. 1
Risk of injuries on mechanical parts!
If not observed, slight to medium injuries, especially to the hands,
can occur.
Observe the notes concerning the C-arm carbon fiber structure.
Measurement A-C: At the handgrip of the horizontal unit, the measuring
points move to 30, 170 and 200 mm
Measurement C-A: At the railing near the image intensifier, the measur-
ing points move to 30, 170 and 200 mm
A B C
Attachment points - spring scale
WARNING