Page 236
MultiRanger 100/200 – INSTRUCTION MANUAL
7ML19981FB05
mmmm
m
App
e
ndix C:
Tr
oubleshoot
ing
larger diameter or shorter length nozzle, bevel the inside of the bottom end, or cut the
open end of the nozzle at a 45° angle.
See the transducer manual for complete mounting instructions.
For ST-series and XPS-10 transducers use the plastic conduit / flange adapter supplied
with the unit.
If the mounting hardware is over tightened, loosen it. Over tightening changes the
resonance characteristics of the transducer and can cause problems.
Set the MultiRanger to Ignore the Bad Echo
If the preceding remedies have not fixed the problem, the false echo has to be ignored.
If the Echo is Close to the Transducer
If there is a static, incorrect, high level reading from the MultiRanger there is probably
something reflecting a strong echo back to the transducer. If the material level never
reaches that point extend the Near Blanking (P800) to a distance to just past the
obstruction.
Adjust the TVT to Ignore the Echo
Use Auto False Echo Suppression (P837-P839) to automatically shape around false
echoes.
Wrong Reading
If the Reading is erratic, or jumps to some incorrect value periodically, ensure the:
1.
Surface monitored is not beyond the MultiRanger’s programmed range or the
transducer’s maximum range.
2.
Material is not falling into the transducer’s acoustic beam.
3.
Material is not inside the blanking distance of the transducer.
Types of Wrong Readings
If a periodic wrong Reading is always the same value, see
Fixed Reading
.
If the wrong Reading is random, ensure the material surface to transducer distance is
less than the Empty value entered plus 20%. If the material/object monitored is outside
this distance, increase Range Extension (P801) as required. This error is most common in
OCM applications using weirs.
Liquid Splashing
If the material monitored is a liquid, check for splashing in the vessel. Enter a lower
Maximum Process Speed (P003) value to stabilize the Reading, or install a stilling well.
(Contact Siemens Milltronics or your local distributor.)
Adjust the Echo Algorithm
Use Dolphin Plus to view live sonic profiles and make adjustments to the P820 Algorithm
parameter. See P820 on page 204 for details.
If the “Area” algorithm is used and narrow noise spikes are evident on the (long shot)
Echo Profile, turn the Spike Filter (P821) on and/or widen the Narrow Echo Filter (P822).
Also, if the true echo has jagged peaks, use Reform Echo (P823).
Summary of Contents for MAG 1100
Page 1: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 47: ......
Page 76: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 140: ...65 SITRANS F M 8 Commissioning 8 3 2 MAG 5000 CT and MAG 6000 CT Menu overview ...
Page 164: ...Tab 36 Siemens MultiRanger 100 ...
Page 165: ...Instruction Manual April 2006 100 200 multiranger ...
Page 396: ...Page 222 MultiRanger 100 200 INSTRUCTION MANUAL 7ML19981FB05 mmmmm Parameters ...
Page 445: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 446: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 454: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 462: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 470: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 478: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 486: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 494: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 502: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 510: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 518: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 526: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 534: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 539: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 540: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 550: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...
Page 555: ...Tab 37 Siemens XRS 5 Ultrasonic Level Sensors ...
Page 556: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 557: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 565: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 573: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 581: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 589: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 597: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 605: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 613: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 621: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 629: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 637: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 645: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 650: ...Tab 38 Siemens XPS XCT Transducers ...
Page 651: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 652: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 662: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...