81
SITRANS F M
8. Commissioning
Error system
The transmitter system is equipped with an error and status log system with 4 groups of information.
•
(I) Information without a functional error involved
•
(W) Warnings which may cause malfunction in the application. The cause of the error may
disappear on its own
•
(P) Permanent errors which may cause malfunction in the application. The error requires an
operator
•
(F) Fatal error which is essential for the operation of the flowmeter
2 menus are available in service and operator menus for registration of information and errors
•
Error pending
•
Status log
Error pending
The first 9 standing errors are stored in “error pending”. When an error is removed it is removed from
“error pending”.
The acceptance level for “error pending” can be individually configured to a particular application.
The acceptance level is set in the “basic settings” in the transmitter setup menu.
Acceptance levels
The following three acceptance levels are selectable
•
Fatal error: Fatal errors are registered as errors
•
Permanent errors: Permanent and fatal errors are registered as errors
•
Warning (Default value): Warnings, permanent and fatal errors are registered as errors
Error information is displayed in title and subtitle lines, see display layout. Title line will show time
since occurrence of error in days, hours and minutes. Subtitle line will flash between an error text
and a remedy text. Error text will indicate type of error (I, W, P or F), error number, and error text.
Remedy text will inform operator of action to take to remove error.
Status Log
Like “error pending” except that information, warnings, permanent and fatal errors is always stored
in the “status log”. The “status log” stores the latest 9 messages received/registered during the last
180 days.
Alarm field
The alarm field on the display will always flash when an error is pending.
Error output
The digital and relay output can individually be activated error by error (error level). The relay output
is default selected to error level. An output can also be selected to activate on a single error number.
The alarm field, error output and error pending always operate together.
The analog output turns to a 1 mA level when in the 4-20 mA mode.
Operator menu
Error pending and status log are as default enabled in the operator menu.
8.7.1
Error handling
3HQGLQJ
(UURUWH[W
Summary of Contents for MAG 1100
Page 1: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 47: ......
Page 76: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 140: ...65 SITRANS F M 8 Commissioning 8 3 2 MAG 5000 CT and MAG 6000 CT Menu overview ...
Page 164: ...Tab 36 Siemens MultiRanger 100 ...
Page 165: ...Instruction Manual April 2006 100 200 multiranger ...
Page 396: ...Page 222 MultiRanger 100 200 INSTRUCTION MANUAL 7ML19981FB05 mmmmm Parameters ...
Page 445: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 446: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 454: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 462: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 470: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 478: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 486: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 494: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 502: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 510: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 518: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 526: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 534: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 539: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 540: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 550: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...
Page 555: ...Tab 37 Siemens XRS 5 Ultrasonic Level Sensors ...
Page 556: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 557: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 565: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 573: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 581: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 589: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 597: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 605: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 613: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 621: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 629: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 637: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 645: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 650: ...Tab 38 Siemens XPS XCT Transducers ...
Page 651: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 652: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 662: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...