29
SITRANS F M
3. Project guidance
The Siemens Flow Instruments cleaning unit can be used with MAG 5000 or 6000 in 19" insert
non-
CT
version.
The cleaning unit can be used in applications where the liner material and subsequently the
electrodes may be coated with deposits. If the coating is electrically insulating, the electrode signal
will be reduced. If the coating is electrically conductive, the electrode signal will be partly short-
circuited and in both cases the accuracy of the meter will decrease (dependent on the type and
thickness of the coating).
Note
Not to be used with sensors with
Hastelloy or Tantalum
electrodes.
The cleaning unit cleans the electrodes electro-chemically by applying a voltage to the electrodes
for approx. 60 sec. While cleaning, the transmitter stores and holds the latest measured flow reading
on the display and also the signal outputs. After an additional pausing period of 60 sec. the flowmeter
resumes normal measurement and the cleaning is now completed.
The relay in the transmitter activates the cleaning cycle. In the relay output menu (under cleaning)
the cleaning interval can be set between 1 hour and 24 hours.
Cleaning should only take place with liquid in the pipe. This can be detected via the empty pipe
detection function. It is therefore recommended to select “empty pipe detection” ON when using the
cleaning unit.
The cleaning sequence can also be controlled manually through the electrical input of the
transmitter. Before this is done, ensure that the measuring pipe is full.
Mode of operation
3.4
Cleaning unit
AC-cleaning
DC-cleaning
AC-cleaning is used to remove fatty deposits on the electrodes. These fatty deposits are seen in
wastewater applications, in abbatoirs and water applications with oil residuals. During the cleaning
process, the surface of the electrodes get warmer, which tends to soften grease particles and the
gas bubbles generated mechanically lift deposits away from the surface of the electrodes.
DC-cleaning is used to eliminate electrically conductive deposits in the measuring pipe influencing
the measuring accuracy.
Particularly in district heating applications an electrically conductive deposit (magnetite) may occur
and short-circuit the electrode signal. In this case the accuracy of the meter decreases and the
signal/noise conditions of the meter become inferior. The problem only arises if the conductivity of
the water is less than approx. 250
μ
S/cm.
During DC-cleaning, electrolysis takes place where the flow of electrons removes the particle
deposits from the electrode area.
Summary of Contents for MAG 1100
Page 1: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 47: ......
Page 76: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Page 140: ...65 SITRANS F M 8 Commissioning 8 3 2 MAG 5000 CT and MAG 6000 CT Menu overview ...
Page 164: ...Tab 36 Siemens MultiRanger 100 ...
Page 165: ...Instruction Manual April 2006 100 200 multiranger ...
Page 396: ...Page 222 MultiRanger 100 200 INSTRUCTION MANUAL 7ML19981FB05 mmmmm Parameters ...
Page 445: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 446: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 454: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 462: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 470: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 478: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 486: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 494: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 502: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 510: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 518: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 526: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 534: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 539: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 540: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 550: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...
Page 555: ...Tab 37 Siemens XRS 5 Ultrasonic Level Sensors ...
Page 556: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Page 557: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 565: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Page 573: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Page 581: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Page 589: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Page 597: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Page 605: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Page 613: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Page 621: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Page 629: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Page 637: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Page 645: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Page 650: ...Tab 38 Siemens XPS XCT Transducers ...
Page 651: ...Operation Manual February 2005 ECHOMAX xps xct ...
Page 652: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Page 662: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...