REMOTE INTERFACE
22/11/2016 Version No. 002
245
Note
Frequency and amplitude cannot be used with the DC function;
however a value (or DEFault) must be specified. The values are
remembered for the next function used.
Syntax
SOURce[1]:APPLy:USER [<frequency> [,<amplitude> [,<offset>] ]]
Parameter
<frequency>
1
μHz~100MHz
<amplitude>
0~10V (50Ω)
<offset>
0~5V (50Ω)
Example
SOUR1:APPL:USER
SOURce[1]:APPLy?
Source Specific
Command
Description
Outputs a string with the current settings.
Note
The string can be passed back appended to the Apply Command.
Syntax
SOURce[1]:APPLy?
Return Parameter
<string>
Function, frequency, amplitude,
offset
Example
SOUR1:APPL?
SIN +5.000000003,+3.0000E+00,-2.50E+00
Returns a string with the current function and parameters, Sine,
5kHz, 3 Vpp, -2.5V offset.
Summary of Contents for AFG-31000 Series
Page 1: ...Instruction Manual AFG 31000 Series Arbitrary Function Generator EN FR DE IT SP ...
Page 2: ......
Page 80: ...AFG 31000 Series Instruction Manual English 78 22 11 2016 Version No 002 ...
Page 112: ...AFG 31000 Series Instruction Manual English 110 22 11 2016 Version No 002 ...
Page 139: ...MODULATION 22 11 2016 Version No 002 137 ...
Page 149: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 147 ...
Page 157: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 155 ...
Page 167: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 165 ...
Page 178: ...AFG 31000 Series Instruction Manual English 176 22 11 2016 Version No 002 ...
Page 181: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 179 ...
Page 183: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 181 ...
Page 185: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 183 ...
Page 210: ...AFG 31000 Series Instruction Manual English 208 22 11 2016 Version No 002 ...
Page 219: ...REMOTE INTERFACE 22 11 2016 Version No 002 217 ...
Page 339: ...REMOTE INTERFACE 22 11 2016 Version No 002 337 ...
Page 360: ...AFG 31000 Series Instruction Manual English 358 22 11 2016 Version No 002 Warning symbol 4 ...