REMOTE INTERFACE
22/11/2016 Version No. 002
289
Description
Turns the trigger out signal on or off from the Trig out terminal on the
rear panel. When set to on, a trigger signal (TTL) is output at the start
of each pulse. The default is setting is off.
Syntax
OUTPut[1]:TRIGger {OFF|ON}
Example
OUT OUTP1:TRIG ON
Enables the Trig out signal.
Query Syntax
OUTPut[1]:TRIGger?
Return Parameter
0
Disabled
1
Enable
Example
OUTP1:TRIG?
1
The Trig out signal is enabled.
SOURce[1]:MARKer:FREQuency
Source Specific
Command
Description
Sets or queries the marker frequency. The default marker frequency
is 500 Hz. The marker frequency is used to output a SYNC signal
from the SYNC terminal on the front panel. The SYNC signal goes
logically high at the start of each sweep and goes low at the marker
frequency.
Note
The marker frequency must be between the start and stop
frequencies. If the marker frequency is set to a value that is out of the
range, the marker frequency will be set to the center frequency and a
“settings conflict” error will be generated.
Summary of Contents for AFG-31000 Series
Page 1: ...Instruction Manual AFG 31000 Series Arbitrary Function Generator EN FR DE IT SP ...
Page 2: ......
Page 80: ...AFG 31000 Series Instruction Manual English 78 22 11 2016 Version No 002 ...
Page 112: ...AFG 31000 Series Instruction Manual English 110 22 11 2016 Version No 002 ...
Page 139: ...MODULATION 22 11 2016 Version No 002 137 ...
Page 149: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 147 ...
Page 157: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 155 ...
Page 167: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 165 ...
Page 178: ...AFG 31000 Series Instruction Manual English 176 22 11 2016 Version No 002 ...
Page 181: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 179 ...
Page 183: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 181 ...
Page 185: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 183 ...
Page 210: ...AFG 31000 Series Instruction Manual English 208 22 11 2016 Version No 002 ...
Page 219: ...REMOTE INTERFACE 22 11 2016 Version No 002 217 ...
Page 339: ...REMOTE INTERFACE 22 11 2016 Version No 002 337 ...
Page 360: ...AFG 31000 Series Instruction Manual English 358 22 11 2016 Version No 002 Warning symbol 4 ...