AFG-31000 Series Instruction Manual/English
126
22/11/2016 Version No. 002
4. Press F2~F6 to choose the frequency
unit.
uHz
F 2
~
MHz
F 6
Range
Frequency
2mHz~80MHz(31081)/
50MHz(31051)
Frequency
– Ramp
2mHz~1MHz
Default
1kHz
Note
Waveform frequency and burst period are not the same. The burst
period is the time between the bursts in N-Cycle mode.
Burst Cycle/Burst Count
The burst cycle (burst count) is used to define the number of cycles that are output for a burst
waveform. Burst cycle is only used with N-cycle mode (internal, external or manual source).
The default burst cycle is 1.
Panel Operation
1. Press the Burst key.
Burst
2. Press F1 (N Cycle).
N Cycle
F 1
3. Press F1 (Cycles).
Cycles
F 1
4. The Cycles parameter will become highlighted in the Waveform
Display area.
Summary of Contents for AFG-31000 Series
Page 1: ...Instruction Manual AFG 31000 Series Arbitrary Function Generator EN FR DE IT SP ...
Page 2: ......
Page 80: ...AFG 31000 Series Instruction Manual English 78 22 11 2016 Version No 002 ...
Page 112: ...AFG 31000 Series Instruction Manual English 110 22 11 2016 Version No 002 ...
Page 139: ...MODULATION 22 11 2016 Version No 002 137 ...
Page 149: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 147 ...
Page 157: ...SECONDARY SYSTEM FUNCTION SETTINGS 22 11 2016 Version No 002 155 ...
Page 167: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 165 ...
Page 178: ...AFG 31000 Series Instruction Manual English 176 22 11 2016 Version No 002 ...
Page 181: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 179 ...
Page 183: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 181 ...
Page 185: ...ARBITRARY WAVEFORMS 22 11 2016 Version No 002 183 ...
Page 210: ...AFG 31000 Series Instruction Manual English 208 22 11 2016 Version No 002 ...
Page 219: ...REMOTE INTERFACE 22 11 2016 Version No 002 217 ...
Page 339: ...REMOTE INTERFACE 22 11 2016 Version No 002 337 ...
Page 360: ...AFG 31000 Series Instruction Manual English 358 22 11 2016 Version No 002 Warning symbol 4 ...