<Power>
The number of results n and the test point separation depend on
the resolution bandwidth, see table below. The value in the mid-
dle of each result array corresponds to the carrier frequency.
For
RBW100
and
RBW1000
results are only available for frequen-
cies, for which limits using these RBW have been defined. For
other frequencies INV is returned.
Range:
-120 dBm to 55 dBm
Default unit: dBm
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.20
The following table lists the number of results n, the test point separation and the fre-
quency ranges depending on the resolution bandwidth.
Table 6-1: <Power> results depending on RBW
RBW
Results n
Separation
Frequency Range
RBW30
4799
15 kHz
-35.985 MHz to +35.985 MHz
RBW100
1599
45 kHz
-35.955 MHz to +35.955 MHz
RBW1000
799
90 kHz
-35.910 MHz to +35.910 MHz
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
6.3.16 I/Q Constellation Results (Traces)
The following command returns the results in the I/Q constellation diagram.
FETCh:LTE:MEAS<i>:ENB:MEValuation:TRACe:IQ?
Returns the results in the I/Q constellation diagram.
See also
chapter 3.8.7, "Detailed Views: I/Q Constellation Diagram"
Return values:
<Reliability>
<I_Phase_1>
<Q_Phase_1> ...
<I_Phase_n>
<Q_Phase_n>
n normalized I and Q amplitudes, one value per modulation sym-
bol
Range:
-2 to 2
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.20
Manual operation:
See
Multi Evaluation Measurement Commands