GUI Reference
R&S
®
CMW-KN5xx
44
User Manual 1176.7690.02 ─ 05
External trigger signal fed in via TRIG A or TRIG B on the rear panel of the instru-
ment.
Remote command:
TRIGger:LTE:MEAS<i>:ENB:MEValuation:SOURce
TRIGger:LTE:MEAS<i>:ENB:MEValuation:CATalog:SOURce?
Trigger Slope
Qualifies whether the trigger event is generated at the rising or at the falling edge of
the trigger pulse. This setting has no influence on "Free Run" measurements and for
evaluation of trigger pulses provided by other firmware applications.
Remote command:
TRIGger:LTE:MEAS<i>:ENB:MEValuation:SLOPe
Trigger Threshold
Defines the input signal power where the trigger condition is satisfied and a trigger
event is generated. The trigger threshold is valid for power trigger sources. It is a dB
value, relative to the reference level minus the external attenuation (<Ref. Level>
–
<External Attenuation (Input)>
–
<Frequency Dependent External Attenuation>). If the
reference level is set to the actual maximum output power of the DUT, and the external
attenuation settings are in accordance with the test setup, then the trigger threshold is
referenced to the actual maximum RF input power at the R&S
CMW.
A low threshold may be required to ensure that the R&S
CMW can always detect the
input signal. A higher threshold can prevent unintended trigger events.
Remote command:
TRIGger:LTE:MEAS<i>:ENB:MEValuation:THReshold
Trigger Delay
Defines a time delaying the start of the measurement relative to the trigger event. This
setting has no influence on "Free Run" measurements.
Remote command:
TRIGger:LTE:MEAS<i>:ENB:MEValuation:DELay
Trigger Timeout
Sets a time after which an initiated measurement must have received a trigger event. If
no trigger event is received, a trigger timeout is indicated in manual operation mode. In
remote control mode the measurement is automatically stopped. The parameter can
be disabled so that no timeout occurs.
This setting has no influence on "Free Run" measurements.
Remote command:
TRIGger:LTE:MEAS<i>:ENB:MEValuation:TOUT
Min Trigger Gap
Defines a minimum duration of the power-down periods (gaps) between two triggered
power pulses. This setting is valid for an "(IF) Power" trigger source.
Parameters and Settings