GUI Reference
R&S
®
CMW-KN5xx
42
User Manual 1176.7690.02 ─ 05
If the checkbox is enabled, the configured ρ
A
value overrides the value of the selected
"E-UTRA Test Model".
Remote command:
CONFigure:LTE:MEAS<i>:ENB:MEValuation:MODulation:PDSCh:RHOA
4.2.4 Spectrum Settings
The following parameters configure the spectrum settings of the measurement. They
are contained in section "Measurement Control".
Fig. 4-4: Spectrum settings
Emission Mask / ACLR > Statistic Count
...................................................................... 42
........................................................................................ 42
Emission Mask / ACLR > Statistic Count
Defines the number of measurement intervals per measurement cycle (statistics cycle,
single-shot measurement). This value is also relevant for continuous measurements,
because the averaging procedures depend on the statistic count.
For emission mask / ACLR results, the measurement interval is completed when the
R&S
CMW has measured a full subframe.
Remote command:
Emission Mask > Meas Filter
Selects the resolution filter type for filter bandwidths of 100 kHz and 1 MHz (gaussian
or bandpass filter). For 30 kHz filters the type is fixed (gaussian). The filter bandwidth
to be used can be configured per emission mask area as part of the limits.
Remote command:
CONFigure:LTE:MEAS<i>:ENB:MEValuation:SPECtrum:SEMask:MFILter
ACLR > Select ACLR
Specifies which types of adjacent channels shall be evaluated for the ACLR results: E-
UTRA channels with the same bandwidth as the carrier, UTRA channels with 1.28
MHz / 3.84 MHz / 7.68 MHz bandwidth. For each enabled type the 1
st
and 2
nd
adjacent
channels are evaluated.
Parameters and Settings