Statistic Count
Defines the number of measurement intervals per measurement cycle (statistics cycle,
single-shot measurement). This value is also relevant for continuous measurements,
because the averaging procedures depend on the statistic count.
For modulation measurements, the measurement interval is completed when the
R&S
CMW has measured a full subframe, containing 14 OFDM symbols. The mea-
surement always extends over the whole subcarrier range.
The measurement provides independent statistic counts for different results. In single-
shot mode a measurement with a completed statistic count is stopped, even if another
measurement with an incomplete statistic count is still running.
Remote command:
CONFigure:LTE:MEAS<i>:ENB:MEValuation:SCOunt:MODulation
EVM Window Length
Length of the EVM window, defined as a number of samples. The standard (3GPP TS
36.104, Annex E) specifies the window length as a function of the bandwidth and the
cyclic prefix. The R&S
CMW uses the "Current" value. This value is linked to the value
in the table, corresponding to the current cyclic prefix (at present only normal CP) and
bandwidth settings.
The window length defines the two sets of "l / h" modulation quantities in the result
tables; see
chapter 3.5, "Calculation of Modulation Results"
The minimum value of 1 FFT symbol actually corresponds to a window of zero length
so that EVM
h
= EVM
l
.
Remote command:
CONFigure:LTE:MEAS<i>:ENB:MEValuation:MODulation:EWLength
CONFigure:LTE:MEAS<i>:ENB:MEValuation:MODulation:EWLength:
CBANdwidth<Band>
EVM Window Position
Selects which results are evaluated for the I/Q diagram - the results calculated at the
low extremities or at the high extremities of the EVM window; see also
"Calculation of Modulation Results"
Remote command:
CONFigure:LTE:MEAS<i>:ENB:MEValuation:MODulation:EWPosition
PDSCH rhoA
ρ
A
(rhoA) defines the power level of a PDSCH resource element relative to the RS
EPRE. It applies if no RS resource elements are transmitted simultaneously on other
subcarriers.
If the checkbox is disabled, the ρ
A
is determined by the selected "E-UTRA Test Model",
Parameters and Settings