6.2 General Measurement Settings
The commands valid for all LTE eNodeB measurements are divided into the groups lis-
ted below.
6.2.1 Duplex Mode
The following command selects the duplex mode.
CONFigure:LTE:MEAS<i>:ENB:DMODe
<Mode>
Selects the duplex mode of the LTE signal: FDD or TDD.
Parameters:
<Mode>
FDD | TDD
Example:
See
Specifying General and Common Measurement Settings
Firmware/Software:
V3.0.20
Options:
FDD requires R&S CMW-KN500
TDD requires R&S CMW-KN550
Manual operation:
See
6.2.2 Signal Routing
The following commands configure the scenario, select the input path for the measured
signal and define an external attenuation value.
ROUTe:LTE:MEAS<i>:ENB:SCENario:SALone
...................................................................61
ROUTe:LTE:MEAS<i>:ENB:SCENario?
............................................................................. 62
CONFigure:LTE:MEAS<i>:ENB:RFSettings:EATTenuation
.................................................. 62
ROUTe:LTE:MEAS<i>:ENB:SCENario:SALone
<RXConnector>, <RFConverter>
Activates the standalone scenario and selects the RF input path for the measured RF
signal.
For possible connector and converter values, see
chapter 6.1.4, "Values for RF Path
Parameters:
<RXConnector>
RF connector for the input path
<RFConverter>
RX module for the input path
Example:
See
Specifying General and Common Measurement Settings
Firmware/Software:
V3.0.20
General Measurement Settings