Command Reference
R&S
®
CMW-KN5xx
89
User Manual 1176.7690.02 ─ 05
<MagErrLow0>
<MagErrHigh0> ...
<MagErrLow13>
<MagErrHigh13>
28 magnitude error values, at low and high EVM window posi-
tion, OFDM symbol 0 to 13
Range:
0 % to 100 %
Default unit: %
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.20
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
6.3.13 Phase Error Results (Traces)
The following commands return the phase error vs. OFDM symbol results of the multi
evaluation measurement.
FETCh:LTE:MEAS<i>:ENB:MEValuation:PERRor:CURRent?
FETCh:LTE:MEAS<i>:ENB:MEValuation:PERRor:AVERage?
FETCh:LTE:MEAS<i>:ENB:MEValuation:PERRor:MAXimum?
READ:LTE:MEAS<i>:ENB:MEValuation:PERRor:CURRent?
READ:LTE:MEAS<i>:ENB:MEValuation:PERRor:AVERage?
READ:LTE:MEAS<i>:ENB:MEValuation:PERRor:MAXimum?
Returns the values of the phase error bar graphs for the OFDM symbols in the mea-
sured subframe. The results of the current, average and maximum bar graphs can be
retrieved.
See also
chapter 3.8.2, "Detailed Views: EVM, Magnitude Error, Phase Error"
Return values:
<Reliability>
<PhaseErrLow0>
<PhaseErrHigh0> ...
<PhaseErrLow13>
<PhaseErrHigh13>
28 phase error values, at low and high EVM window position,
OFDM symbol 0 to 13
Range:
0 deg to 180 deg
Default unit: deg
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.20
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
Multi Evaluation Measurement Commands