![Rohde & Schwarz R&S CMW-KN5 Series User Manual Download Page 81](http://html.mh-extra.com/html/rohde-and-schwarz/rands-cmw-kn5-series/rands-cmw-kn5-series_user-manual_1477861081.webp)
Command Reference
R&S
®
CMW-KN5xx
81
User Manual 1176.7690.02 ─ 05
Parameters:
<RMS>
Range:
0 deg to 180 deg
*RST:
18.5 deg, OFF
Default unit: deg
Additional parameters: OFF | ON (disables | enables the limit
check)
<Peak>
Range:
0 deg to 180 deg
*RST:
35 deg, OFF
Default unit: deg
Additional parameters: OFF | ON (disables | enables the limit
check)
Example:
See
Firmware/Software:
V3.0.20
CONFigure:LTE:MEAS<i>:ENB:MEValuation:LIMit:QPSK:FERRor
<FrequencyError>
Defines an upper limit for the carrier frequency error (QPSK modulation).
Parameters:
<FrequencyError>
Range:
0 ppm to 1 ppm
*RST:
0.25 ppm, ON
Default unit: ppm
Additional parameters: OFF | ON (disables | enables the limit
check)
Example:
See
Firmware/Software:
V3.0.20
CONFigure:LTE:MEAS<i>:ENB:MEValuation:LIMit:QPSK:IQOFfset
<Offset>
Defines an upper limit for the I/Q origin offset (QPSK modulation).
Parameters:
<Offset>
I/Q origin offset limit for TX power
Range:
-256 dBc to 256 dBc
*RST:
-24.2 dBc, OFF
Default unit: dBc
Additional parameters: OFF | ON (disables | enables the limit
check)
Example:
See
Firmware/Software:
V3.0.20
Multi Evaluation Measurement Commands
深圳德标仪器
135-1095-0799