![Rohde & Schwarz EX-IQ-BOX User Manual Download Page 371](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982371.webp)
Interfaces
R&S
®
EX-IQ-BOX
356
User Manual 1409.5570.32 ─ 05
Information on Z-DOK-Adapter Board Connector
The described pin assignment applies to applications with the R&S
EX-IQ-BOX controlled
by R&S
DigIConf. If an R&S Instrument controls the R&S
EX-IQ-BOX directly, the pinout
varies. For information on the differences, refer to
chapter 10.1.3, "Advanced pinout with
Find detailed information on the PC Board Footprint, the Housing and the Block Diagram
Mating Face Configuration in the customer drawing C-6367557, attached to the operating
manual.
Table 10-1: Pinout Z-DOK-Adapter Board Connector (User Interface)
Signal Name
Pin
Dir
Description
BO_TYPE0
A2
I
breakout board identification or
EEPROM
BO_TYPE1
A1
I
BO_SENS
Z3, X3, Y3, W3
breakout board sense
UI_I_N/P0 ... UI_I_N/P17
F1/F2 ... C9/C10
I/O
I data (real part)
UI_Q_N/P0 ... UI_Q_N/P17
C13/C14 ... D23/D24
I/O
Q data (imaginary part)
UI_VALID_N/P
F13/F14
I/O
indicates valid data blocks
Note:
This signal must be active in
receiver mode.
D_CLK_UIN_N/P
F25/F26
I
interface clock input.
Note:
This clock signal is required as
a reference for PLL. Therefore, it
must be active continuously.
D_CLK_UOUT_N/P
A13/A14
O
interface clock output
UI_GP_N/P0 ... UI_GP_N/P5
[1]
F11/F12 ... A25/26
I/O
additional protocol dependent con-
trol / signaling. If not needed the sig-
nal can be unconnected.
UI_AUXIO_N/P0 ... UI_AUXIO_N/P3 A23/A24 ... D11/D12
I/O
additional auxiliary signals
TX_N_A/TX_P_A
A27/A28
O
high speed serial transmitter output.
These signals are used by standar-
dized protocols.
TX_N_B/TX_P_B
F27/F28
O
TX_N_C/TX_P_C
D2/D1
O
RX_N_A/RX_P_A
D28/D27
I
high speed serial receiver input.
These signals are used by standar-
dized protocols.
RX_N_B/RX_P_B
C28/C27
I
RX_N_C/RX_P_C
C26/C25
I
[1]
GP0 ... GP5 correspond to the General Purpose signals 0...5 as they are named in
other R&S instruments and R&S devices.
The N/P notation applies to differential interface standards (LVDS). For single ended